Specifications |
| Oxygen Gas(at 1 atmosphere of total pressure) | Dissolved Oxygen in Water(at 25 ℃) |
Dynamic range: | 0-100 mole%; 0-760 mm Hg partial pressure | 0-40.7 ppm; 0-760 mm Hg partial pressure |
Resolution: | 0.06% at 0% Oxygen 0.3% at 20% Oxygen 5% at 100% Oxygen | 0.02 ppm at 0 ppm Oxygen 0.12 ppm at 8.5 ppm Oxygen 2.04 ppm at 40.7 ppm Oxygen |
Stability: | ±0.03 mole%; ±0.23 mm Hg | ±0.012 ppm; ±0.23 mm Hg |
Minimum detection: | 0.1% mole%; 0.76 mm Hg | 0.04 ppm |
Accuracy(0-25 mole%): | Linear(2-point Minimum Calibration):3% Polynomial(3-point Minimum Calibration):2% | Linear(2-point Minimum Calibration):3% Polynomial(3-point Minimum Calibration):2% |
Accuracy(0-100 mole%): | Linear(2-point Minimum Calibration):30% Polynomial(3-point Minimum Calibration):6% | Linear(2-point Minimum Calibration):30% Polynomial(3-point Minimum Calibration):6% |
Response time: | < 1 second(no overcoat necessary) | < 1 second(no overcoat necessary) |
Operating temperature: | 0 ℃ to 50 ℃ | 0 ℃ to 50 ℃ |
Sterilization: | Gamma radiation or sodium hypochlorite(bleach or Clorox) | Gamma radiation or sodium hypochlorite(bleach or Clorox) |
Probe assembly: | 1000 um fiber in stainless steel ferrule | 1000 um fiber in stainless steel ferrule |
Probe dimensions: | 1.587 mm OD × 152.4 mm length | 1.587 mm OD × 152.4 mm length |