类型:数字示波器品牌:LeCroy型号:WaveRunner 204MXi-A
通道数:4显示模式:Windows lcd最大输入:600(V)
量程:10(V/div) 电源电压:220(V) 外形尺寸:10.25 x 13.4x 6(mm)
重量:6.95(kg)
Vertical System
Model
WaveRunner 204MXi-A
WaveRunner 104MXi-AWaveRunner 64MXi-AWaveRunner 44MXi-A
Nominal Analog Bandwidth @ Ω 50 10 mV-1 V/div
2 GHz
1 GHz600 MHz400 MHz
Rise Time (typical)
180 ps
300 ps500 ps875 ps
Input Channels
4
444
Bandwidth Limiters
20 MHz; 200 MHz
Input Impedance
1 MΩ || 20 pF or 50 Ω
1 MΩ || 20 pF or 50 Ω1 MΩ || 16 pF or 50 Ω1 MΩ || 16 pF or 50 Ω
Input Coupling
50 Ω: DC, 1 MΩ: AC, DC, GND
Maximum Input Voltage
50 Ω: 5 Vrms, 1 MΩ: 400 V max. (DC + Peak AC = 5 kHz)
Input Connector
ProBus/BNC
Vertical Resolution
8 bits; up to 11 with enhanced resolution (ERES)
Sensitivity
50 Ω: 2 mV/div? V/div fully variable; 1 MΩ: 2 mV?0 V/div fully variable
Offset Range
?00 mv@2?.95 mv/div, ? V@5?9 mv/div, ?0 V @ 100 mv? V/div, ?00V @ 1.02?0 V/div; 1 M O: ?00 mv @ 2?.95 mv/div, ? V @ 5?9 mv/div, ?0V @ 100 mv? V/div, ?00V @ 1.02?0 V/div

Main Specifications
Probes
One PP007 per channel standard; Optional passive and active probes available.

Timebase System
Roll Mode
User selectable at <= 500 ms/div and <= 100 kS/s

Horizontal System
Timebases
Internal timebase common to all input channels; an external clock may be applied at the auxiliary input
Time/Division Range
Real time: 200 ps/div?0 s/div, RIS mode: 200 ps/div to 10 ns/div, Roll mode: up to 1,000 s/div
Clock Accuracy
<= 5 ppm @ 25 蚓 (typical) (<= 10 ppm @ 5?0 蚓)
Trigger and Interpolator Jitter
Jitter <=3 ps rms (typical)
Sample Rate and Delay Time Accuracy
Equal to Clock Accuracy
Channel-Channel Deskew Range
? x time/div setting, 100 ms max., each channel
External Sample Clock
DC to 600 MHz; (DC to 1 GHz for 104Xi and 204Xi) 50 Ω, (limited BW in 1 MΩ), BNC input, limited to 2 Ch operation (1 Ch in 62Xi-A), (minimum rise time and amplitude requirements apply at low frequencies)

Acquisition System
Single-Shot Sample Rate/Ch
5 GS/s
Interleaved Sample Rate (2 Ch)
10 GS/s
Random Interleaved Sampling (RIS)
200 GS/s
RIS Mode
User selectable from 100 ps/div to 10 ns/div
Maximum Trigger Rate
1,250,000 waveforms/second
Acquisition Memory Options
Max. Acquisition Points (4 Ch/2 Ch, 2 Ch/1 Ch in 62Xi-A) Segments (Sequence Mode)
Sequence Time Stamp Resolution
1 ns
Standard
12.5M/25M 10,000

Acquisition Processing
Averaging
Summed and continuous averaging to 1 million sweeps
Enhanced Resolution (ERES)
From 8.5 to 11 bits vertical resolution
Envelope (Extrema)
Envelope, floor, or roof for up to 1 million sweeps
Interpolation
Linear or (Sinx)/x
Time Resolution (min, Single-shot)
100 ps (10 GS/s)

Triggering System
Trigger Modes
Normal, Auto, Single, Stop
Sources
Any input channel, External, Ext/10, or Line; slope and level unique to each source, except Line
Coupling
DC, AC (typically 7.5 Hz), HF Reject, LF Reject
Pre-trigger Delay
0?00% of memory size (adjustable in 1% increments, or 100 ns)
Post-trigger Delay
Up to 10,000 divisions in real time mode, limited at slower time/div settings in roll mode
Hold-off by Time or Events
1 ns to 20 s or 1 to 1,000,000,000 events
Internal Trigger Range
?.1 div from center (typical)
Trigger Sensitivity with Edge Trigger (Ch 1?)
2 div @ < 2 GHz 1 div @ < 200 MHz
External Trigger Input Range
EXT/10 ? V; EXT ?00 mV
Max. Trigger Frequency, SMART Trigger
2 GHz @ = 10 mV
Minimum Time Between Sequential Segments Sequential Segments
800 ns

Basic Triggers
Edge
Triggers when signal meets slope (positive, negative, or Window) and level condition.
TV-Composite Video
Triggers selectable fields (1, 2, 4, or 8), Positive or Negative slope, or Line (up to 1500), for NTSC, PAL, SECAM, or non-standard video (up to 1500 lines).

SMART Triggers
State or Edge Qualified
Triggers on any input source only if a defined state or edge occurred on another input source. Delay between sources is selectable by time or events.
Dropout
Triggers if signal drops out for longer than selected time between 1 ns and 20 s.
Pattern
Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input ?2 Ch+EXT on WaveRunner 62Xi-A). Each source can be high, low, or don care. The High and Low level can be selected independently. Triggers at start or end of the pa

SMART Triggers with Exclusion Technology
Glitch and Pulse Width
Triggers on positive or negative glitches with widths selectable from 500 ps to 20 s or on intermittent faults (subject to bandwidth limit of oscilloscope).
Signal or Pattern Interval
Triggers on intervals selectable between 1 ns and 20 s.
Timeout (State/Edge Qualified)
Triggers on any source if a given state (or transition edge) has occurred on another source. Delay between sources is 1 ns to 20 s, or 1 to 99,999,999 events.
Runt
Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 s.
Slew Rate
Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 s.
Exclusion Triggering
Trigger on intermittent faults by specifying the normal width or period.

Color Waveform Display
Type
Color 10.4" flat-panel TFT-LCD with high resolution touch screen
Resolution
SVGA; 800 x 600 pixels; maximum external monitor output resolution of 2048 x 1536 pixels
Number of traces
Display a maximum of 8 traces. Simultaneously display channel, zoom, memory, and math traces.
Grid Styles
Auto, Single, Dual, Quad, Octal, XY, Single + XY, Dual + XY
Waveform Representation
Sample dots joined or dots only in real-time mode

Internal Waveform Memory
Internal Waveform Memory
M1, M2, M3, M4 Internal Waveform Memory (store full-length waveform with 16 bits/data point) or store to any number of files limited only by data storage media.

WaveStream?Fast Viewing Mode
Intensity
256 Intensity Levels, 1?00% adjustable via front panel control
Number of Channels
up to 4 simultaneously
Max Sampling Rate
5 GS/s (10 GS/s for WaveRunner 62Xi-A, 64Xi-A, 104Xi-A, 204Xi-A in interleaved mode)
Waveforms
up to 8000 waveforms/second
Operation
Front panel toggle between normal real-time mode and LeCroy WaveStream Fast Viewing mode

Automatic Setup
Auto Setup
Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals.
Vertical Find Scale
Automatically sets the vertical sensitivity and offset for the selected channels to display a waveform with maximum dynamic range.

Probes
Probe System: ProBus
Automatically detects and supports a variety of compatible probes.
Scale Factors
Automatically or manually selected, depending on probe used

Setup and Waveform Storage
Front Panel and Instrument Status
Store to the internal hard drive, over the network, or to a USB-connected peripheral device.

Zoom Expansion Traces
Zoom Expansion Traces
Display up to 4 Zoom/Math traces with 16 bits/data point

Interface
Remote Control
Via Windows Automation, or via LeCroy Remote Command Set
GPIB Port (optional)
Supports IEEE ?488.2
Ethernet Port
10/100/1000Base-T Ethernet interface (RJ-45 connector)
USB Ports
5 USB 2.0 ports (one on front of instrument) supports Windows-compatible devices.
External Monitor Port
Standard 15-pin D-Type SVGA-compatible DB-15; connect a second monitor to use extended desktop display mode with XGA resolution.
Serial Port
DB-9 RS-232 port (not for remote oscilloscope control)

Auxiliary Input
Signal Types
Selected from External Trigger or External Clock input on front panel
Coupling
50 Ω: DC, 1 MΩ: AC, DC, GND
Maximum Input Voltage
50 Ω: 5 Vrms, 1 MΩ: 250 V max. (DC + Peak AC = 5 kHz)

Auxiliary Output
Signal Types
Trigger Enabled, Trigger Output. Pass/Fail, or Off
Output Level
TTL, ?.3 V
Calibrator Signal
Output available on front panel connector provides a variety of signals for probe calibration and compensation.

General
Auto Calibration
Ensures specified DC and timing accuracy is maintained for 1 year minimum.
Power Requirements
90?64 Vrms at 50/60 Hz; 115 Vrms (?0%) at 400 Hz, Automatic AC Voltage Selection Installation Category: 300V CAT II; Max. Power Consumption: 340 VA/340 W; 290 VA/290 W for WaveRunner 62Xi-A

Environmental and Safety
Temperature (Operating)
+5 蚓 to +40 蚓
Temperature (Non-Operating)
-20 蚓 to +60 蚓
Humidity (Operating)
Maximum relative humidity 80% for temperatures up to 31 蚓 decreasing linearly to 50% relative humidity at 40 蚓
Humidity (Non-Operating)
5% to 95% RH (non-condensing) as tested per MIL-PRF-28800F
Altitude (Operating)
Operating Up to 2,000 m
Altitude (Non-Operating)
Non-Operating 12,190 m

Physical Dimensions
Dimensions (HWD)
260 mm x 340 mm x 152 mm Excluding accessories and projections (10.25" x 13.4" x 6")
Net Weight
6.95 kg. (15.5 lbs.)

Certifications
Certifications
CE Compliant, UL and cUL listed; Conforms to EN 61326, EN 61010-1, UL 61010-1 2nd Edition, and CSA C22.2 No. 61010-1-04.

Warranty and Service
Warranty and Service
3-year warranty; calibration recommended annually. Optional service programs include extended warranty, upgrades, calibration, and customization services.

公司联系方式
  • 东莞市力高大同自动化设备有限公司 [加为商友]
  • 联系人陈从英(女士)  
  • 地区广东-东莞市
  • 地址中国 广东 东莞市 东莞市南城区簪花路华凯活力中心408室
  •   
同类产品

[ 供应搜索 ]  [ ]  [ 告诉好友 ]  [ 打印本文 ]  [ 关闭窗口 ]  [ 返回顶部 ]

CK365测控网郑重提醒:网上过低的价格有可能是虛假价格信息,请买家谨慎对待,谨防价格欺诈行为。投诉或价格欺诈行为,请发送邮件至 1365099215@qq.com